Error Status code: 401
This is to remind you the basic rules of fluorescence lamps and gas lasers:
A multi-photon, or two-photon, microscope uses instrumentation similar to that of a laser scanning confocal microscope: A laser source for sample excitation, a scanhead with galvanometer controlled mirrors (or acousto-optic deflectors) to scan the excitation beam, and photomultiplier tubes to detect fluorescent signals. However, the confocal image differs in that optical sectioning is obtained using excitation in multi-photon microscopy whereas in confocal microscopy it is achieved using the emission pinhole.
The Prairie Ultima was the first commercial system designed to include two sets of conventional galvanometer based scanners, the standard imaging set and the optional uncaging set, for simultaneous imaging and uncaging using two lasers.
Integrated with a custom modified fixed stage microscope, the Ultima is ideal for studies involving the use of tissue slices or in-vivo work on small organisms such as mice.
Location Room :
Manufacturer : Prairie
Model: Ultima
Year Installed: 2008
Keywords:
Still yet to be updated. Soon.
We would like to thank the technical support of IGC's Advanced Imaging Facility (AIF-UIC), which is supported by the national Portuguese funding ref# PPBI-POCI-01-0145-FEDER-022122, co-financed by Lisboa Regional Operational Programme (Lisboa 2020), under the Portugal 2020 Partnership Agreement, through the European Regional Development Fund (FEDER) and Fundação para a Ciência e a Tecnologia (FCT; Portugal).
Configuration |
Value |
Magnification |
Objective type |
NA1 |
Immersion |
Working distance (mm) |
Resolution2 @525nm in XY (µm) |
Resolution @525nm in Z (µm) |
Link |
|
10x |
![]() |
UMPLAN FL |
0.30 |
Water |
3.5 |
1.068 |
1.050 |
|
20x |
![]() |
XLUMPLAN FL N |
1 |
Water |
2 |
0.320 |
0.315 |
|
40x |
![]() |
LUMPLAN FL\IR |
0.80 |
Water |
3.3 |
0.400 |
0.394 |
|
60x |
![]() |
LUMPLAN FI\IR |
0.90 |
Water |
2 |
0.356 |
0.350 |
ref3 |
(1) NA - Numerical Aperture. For more information, follow this link.
(3) This is a similar objective to the one currently in the Prairie, Olympus does not have information for our objective.
The pixel size in microns of your picture is saved automatically. Use LOCI to recognize it with ImageJ.
Check the cooler. The temperature should be around 20 Cº.
Switch the laser from "stand by" to "on" by clockwise turning the key (turn the key to the right).
Turn on the main power switch (white switch on the side of two-photon main tower).
Turn on the computer.
Override the detectors protection.
Switch the detectors protection to the "override" position. Then, put them back in the "protection" position. This step is important to preserve the detectors, also called "PMTs". In this picture, switches are shown in green and red.
Check the filter cubes that you will use and put them in the right positions.
This will be determined according to the wavelengths ("colors") that you want to separate in each PMT. You will probably need the help of UIC members to set up the best filters for your conditions.
This picture shows the 2 upper detectors (PMTs) and the upper filter slot. The filter slot is fixed by magnets.
Keep the filter slot with the little pin at the right side (the red arrow shows the pin).
Every filter cube has a notch. Keep it also at the right side.
Mount the filter cube on top of the filter slot by sliding it over the rail. Note that the pin (filter slot) and notch (cube) are both at the right side!
This should be the final assembly of your filter on top of the filter slot. Any doubts, contact UIC.
Put the whole piece back. Remember: notch and pin to the right side!
Lower filter/detectors chambers
The same mounting steps are required here. Note the red arrow is showing the pin and notch regions. Pin and notch facing the top side!
Change the beam expander switcher to the objective set that you are using.
This will allow the photons coming from the 2-photon laser to fill the back of your objective, optimizing the amount of photons to reach it. Basically, you will switch between 20x or 40/60x.
Check the cable attached to the 4th connector. Change it according to your setup.
There are 3 different cables that can be attached to this connector. By default, the cable for the 4th channel upper detector should be attached.
All the procedures from now on will be done only at the software level. Do not change anything else! Switching buttons, tuning dials, etc. will mess up the microscope configuration, making its use unavailable for you and all the other UIC users until fixing by one of UIC members!
Don't forget that we regularly run theoretical courses, which should give you the basis for understanding what you are doing. If you don't think it is enough, just talk with us for further discussion.
In most of the next pictures, you will find arrows indicating the place to find the command.
Start the "Prairie View" software ("Prairie View.exe" icon).
Find the "2-P Laser" tab and click on "closed" underneath "shutter" box. Then, type the target wavelength and click "enter".
This step will tune the laser to the wavelength needed for your fluorophore. At this point, the laser path will be available to reach the sample, but it will not hit the specimen yet.
This pictures show the location of "2-P Laser" tab, the "close" button that you will need to click and the place to type the wavelength that you will use.
The software does not select/change automatically for every single objective that you are using! You have to select it by clicking on the objective field at Prairie View software, indicated by the green arrow.
Objective installation: by default, the 20x objective will be installed. This objective has a numerical aperture of 1. However, you can use the 40x or 60x objectives despite their numerical aperture of 0.8 and 0.9, respectively. In this case, please ask first to someone at UIC to teach you how to perform this exchange.
Please, KEEP ALL THE MATERIAL NOT IN USE INSIDE THE "4 CHANNEL BOX"! This minimizes the risk of damaging the objectives, dichroics, filters, and also avoid loss of time to find the material for the next users!
Increase the light intensity of the fluorescent lamp, select the correct filter in the fluorescent filter wheel (position 3 for GFP, CFP, etc; position 4 for RFP, Rhodamine, etc) and find your region of interest. To observe the sample at the eye piece, three shutters/switchers have to be opened:
These arrows show exactly how your microscope should be to allow the specimen visualization by the binocular. Most of the time, users cannot see the fluorescence on their tissue because they forget to take the steps mentioned above.
The fluorescent lamp is already turned on, now you must adjust the light intensity.
At the end of your search for the region of interest, take these steps to have the microscope ready for laser scanning:
Now you will need to increase the electrical power (voltage) in the pocket cell and the detectors.
Very important observation: : the PMTs are very sensitive and the room light will burn them. Little by little, if the room lights are turned on at the same time than PMTs, they will lose their capacity to detect dim fluorescence and only very bright signals will be visible. This type of action goes against the 2-photon laws! Therefore, BEFORE turn on the room lights, ALWAYS turn off the PMTs by decreasing the voltage to zero!
Generally a power range among 20% to 30% (cameleon range goes from 0 to 500, where 500 corresponds to 100%) in the pocket cell is more than enough (the higher is the power in the pocket cell, the higher is the probability to cause photodamage!). Usually, the detector power is used at the range of 600 to 800.
Choosing the right channels.
Below you have the location of each channel and the configuration map of all dichroics and filters installed. Using this map you can figure out what will be the channels that you need. As usual, help from UIC members is available.
PMTs and corresponding channels
Now it's time for an important check-point:
Start the live scan of your sample.
You should see the image in the screen. To fine tune your image, adjust the pockel cell and PMTs voltages. You can also change the dwell time. The higher is the dwell time, the better contrast you will have. However, the catch here is that you will also increase the tissue photo damage probability. You can also adjust your gain and offset by clicking on the green tab in the middle of the PMTs tab (in this case, no photo damage problems because you are not dealing with the amount of laser that goes to your sample but improving your image quality only). You can also increase the image resolution by changing the pixelation (for example, from 512x512 to 1024x1024). But, again, the higher is the image resolution, the slower is the image acquisition (and the final file will be very big).
Define your "z-starting" and "z-ending" positions ("how deep you will go").
To define your z-starting position, first find the 3 big knobs on the multi-photon command tower that remotely control the X-Y-Z position of your sample. Turn the Z-knob until you reach the first imaging plane. After that, on the software screen, click the button with black lines followed by one red line. You just select one of your ends. Then, using again the Z-knob, go to the other limit of your z-stack. Once there, click on the button with one red line followed by black lines. You selected the other end of your z-stack.
Acquiring a "z-stack" series ("3D image of your sample").
Select the "Z-Series" tab on your software. There you will find 4 parameters: first step, last step, step size, and number of steps. You will see that you can change up to 3 of these parameters. The other will be automatically selected by the software. Usually, you will only have to change the step size or the number of steps. This is determined according to your previous experience with your samples. Remember: too many step sizes, your file will be huge. Too little, you can lose tracking details between z-stacks.
Select/create the folder to save your data.
After determining the number of z-stacks or the step size between your z-stacks, browse the right folder to save your image. Select the folder or create a new one. IMPORTANT: Always use "D:\All Users" to create your user folder where all the next acquisitions have to be saved.
Acquire z-stack image.
Now you are ready to acquire a static image of your sample (if you don't want to do this, go directly to the next step). Otherwise, just press the button "Start Z-Series" at the lower right side of this tab. The software will acquire all pre-determined z-stacks in this tissue and open a new window with them for your examination. It's already saved on that specified folder and you can exit this new window (now showing your recent acquisition) anytime that you want by selecting "exit" button.
Acquiring a "t-series" image ("how to mount a time-lapse movie of your sample").
If you want to acquire images of your tissue at different time-points, then go to the "T-Series" tab. There, erase all previous program lines already determined by the previous user by clicking on "delete all lines" button. Now, add a new line. The next steps are all related to a "t-series" image and you will find all the places to add your new information in this tab.
In summary, these are the steps described above:
Once you finish your images acquisition, exit the Prairie software and transfer your files to your xserver02 folder, external HD, pendrive, or DVD.
Important Note: It's strictly forbidden to keep the files in the computer HD since it will go against the 2-photon computer laws (after all, the software will run slower, the next user can have problems to save his/her files, and the final outcome is "virtual damage to the software function"). The files will be deleted immediately by UIC members, with no further notice, every Friday.
Finally, to turn off the whole equipment, you will procedure backwards to the turning on procedure:
Turn off the 2-photon command tower by the white switch.
You will hear the sound of all shutters closing.
Turn the key position (on the laser box) to "stand by".
It will start to decrease the laser power and you will hear a noise during this process.