Zeiss High Content Screening | Reservations

  • Zeiss AxioObserver
  • The Zeiss Colibri is an automated research microscope that offers high sensitivity to fluorescence and the option to perform Phase Contraste (Ph). The Solid State Ilumination through LEDs gives a high range of colors and fast aquisition times.

    Wing : Christopher Colombo, P2 Room

    Manufacturer : Zeiss

    Model : Axio Observer Z1

    Year Installed: 2013

    Camera: Hamamatsu Flash Orca 4.0 sCMOS 2048*2048 px

Suggestion for description in "Materials and Methods":

Text under construction.

Suggestion for "Acknowledgments":

We would like to thank the technical support of IGC's Advanced Imaging Facility (AIF-UIC), which is supported by the national Portuguese funding ref# PPBI-POCI-01-0145-FEDER-022122, co-financed by Lisboa Regional Operational Programme (Lisboa 2020), under the Portugal 2020 Partnership Agreement, through the European Regional Development Fund (FEDER) and Fundação para a Ciência e a Tecnologia (FCT; Portugal).

Maintenance log:


Available objectives

Magnification
Objective type
NA
Immersion
Working distance (mm)
Pixel Size (µm)
Pixel Size x1.6 (µm)
Phase Contraste prism
Link
20x
Plan Apochromat
0.8
0
0.55
0.33
0.20
Ph 2
40x
Plan Apochromat
0.95
0
0.25
0.16
0.10
Ph 3
63x
LCI Plan-Neofluar
1.3
Water-Gly*
0.17
0.10
0.06
Ph 3

(*) The 63x objective has a correction ring for different imersion media. It can be used with water or glycerol, but it has to be corrected for which to use. Speak with the UIC personnel if you need to change it.


Filter Sets

Filter Cube
Excitation filter
Dichromatic mirror
Suppression filter
Link
DAPI
BP 377/28
FT 403
BP 464/100
GFP
BP 470/40
FT 495
BP 525/50
dsRed
BP 550/25
FT 570
BP 605/70
Cy5
BP 640/30
FT 660
BP 690/50

Setup

Turn on procedures

  1. Turn on Definite Focus control box (if this is not the first, there might be problems with the this function).
  2. Turn on Colibri control box (LED fluorescent light source).

    • You have to turn on the LEDs you are going to use and set their intensity.
  3. Turn on the microscope control box.
  4. Turn on the microscope (button on the left side of the microscope body).
  5. Turn on the XY-stage control box.
  6. Turn on the camera.
  7. Start the MicroManager software.
  8. Set up your experiments on the microscope.

Turn off procedures

  1. Turn off the microscope control box.
  2. Turn off the camera.
  3. Turn off the Colibri control box.
  4. Turn off the XY-stage control box.
  5. Turn off the Definite Focus control box.
  6. If you used any immersion lenses be sure to clean them.
  7. TAKE OUT EVERYTHING YOU BROUGHT!, eg, pipettes, slides, etc...

Acquisition

  • Grid Colection / Tile

    It is possible to acquire grids and tiles of big samples through Zen software. For a more detailed information, please check the following link or click in the image to the right.

  • Tile Scan

  • Definite Focus

    Definite Focus is Zeiss's version of Autofocus through hardware. This funcionality allows continuous focus of the sample through coverslip position, keeping the sample focused during the whole acquisition with the main objective of preventing drift in Z over long periods of time. For more detailed information, please visit the following link.

  • Zeiss Definite Focus


Extra Info

Widefield Fluorescence

Phase Contrast (Ph)